BRUKER S8 TIGER Series 2 X射線熒光光譜儀 Overview
S8 TIGER Series 2
Thanks to HighSense technology the S8 TIGER Series 2 WDXRF spectrometer delivers for all elements accuracy and precision for industrial quality and process control.
With high resolving WDXRF technology and optimal detection of light, medium and heavy elements based on the new HighSense beam path, the XRF2 mapping tool of the S8 TIGER Series delivers best sensitivity, smallest spot size, and highest resolution for small spot applications.
Ergonomic and failsafe operation are vital for efficiency and best analytical data. Simple, intuitive start of samples are guaranteed with the multilingual TouchControl interface.
HighSense Technology
The S8 TIGER Series 2 comes with the revolutionary HighSense? technologyfor WDXRF. It offers an exceptional linear dynamic range enabling higher precision levels for process and quality control in cement, minerals, mining, metals, glass, and ceramics.
XRF2 Small spot mapping
The S8 TIGER Series 2 features XRF2 small spot mapping which provides detailed maps of the elemental distribution in materials. With a 300?m spot size, XRF2 offers the highest resolution available in a WDXRF instrument.
EZ Ergo - Easy Operation
The S8 TIGER Series 2 comes with intuitive and failsafe operation based on the new multilingual TouchControl interface. With EasyLoad it is designed for high sample throughput. Optimal instrument uptime and low cost of operation are ensured with the unique SampleCare feature - simply EZ Ergo.
BRUKER S8 TIGER Series 2 X射線熒光光譜儀 Technical Details
Systems | S8 TIGER 1 kW 1kW 50 kV max. 50 mA max. | S8 TIGER 3 kW 3 kW 60 kV max. 150 mA max. | S8 TIGER 4 kW 4 kW 60 kV max. 170 mA max. |
X-ray tubes | Rh, 75 μm Be Window | Rh, 75 μm Be Window Rh, 50 μm Be Window, SampleCare protective coating Rh, 28 μm Be Window, SampleCare protective coating Cr, max.3.3 kW | Rh, 75 μm Be Window Rh, 50 μm Be Window, SampleCare protective coating Rh, 28 μm Be Window, SampleCare protective coating Cr, max.3.3 kW |
Collimator | Automatic collimator changer (up to 4) | ||
Analyzer crystals | Automatic crystal changer (up to 8) Included: XS-55, PET, LiF (200) Optional: XS-B, XS-C, XS-N, XS-PET-C, XS-CEM, XS-Ge-C, LiF (220), LiF (420), ADP, Ge, TlAP, InSb, XS-N-HS, XS-100, XS-400, XS-InSb-C | ||
DynaMatch? | Linear intensity range more than 13 million cps | ||
HighSense Detectors | Proportional flow counter and Scintillation Counter: Capacity more than 3 Mio cps, 2 Mio cps linear range | ||
XRF2 | High Resolution and high Intensity Element Mapping with 0.3 mm and 1.25 mm spot (FWHM) | ||
Automated Mask Changer | Automatic masks : closed (SampleCare;) 34 mm, 28 mm, 23 mm, 18 mm, 8 mm (BeamGuide), 5 mm (BeamGuide) | ||
TouchControl? 1) | Integrated touchscreen for easy and intuitive operation | ||
SampleCare? 1) | X-ray tube and goniometer protected by contamination shields Sample and spectrometer chamber separated by programmable vacuum seal | ||
EasyLoad? 1) | Automatic sample recognition Portable sample trays | ||
Power requirements | 208 ? 240 V (1P/3P) 50/60 Hz | 208 V, 60 Hz (1P/3P) 230 V, 50/60 Hz (3P) | |
Compressed air | Not required | ||
Detector gas | P5 gas (5% methane, 95% argon) for flow counter P10 gas (10% methane, 90% argon) for flow counter | ||
External cooling water | No cooling water | Cooling water Water consumption automatically regulated and minimized, short term interruptions are compensated | |
Dimensions (height x width x depth) | 135 cm x 84 cm x 90 cm; 53.1” x 33.1” x 35.4” | 135 cm x 84 cm x 104 cm; 53.1” x 33.1” x 41” | |
Touchscreen: Allows additional width of 49 cm (19.3’’) | |||
446 kg | 476 kg | ||
Quality & safety | DIN EN ISO 9001:2008 CE certified Fully radiation protected system; radiation < 1="" μsv/h="" (bfs="" 09/07="" v=""> |