Salt solutions
Equipped with Rotronic humidity standards, a suitable calibration device and the HW4 software, it is easy to calibrate and adjust probes on-site at your premises. It is also possible to calibrate and adjust probes with the handheld instrument HP32-A (HW4 software is not necessary).
Calibration devices
Rotronic calibration devices are small, airtight chambers that fit Rotronic probes precisely. The lower part of the device consists of a screw-on lid into which the humidity standard is poured on to an absorbent textile pad. The specified humidity is generated in the calibration device after a stabilization period. High humidity values require a longer stabilization period. The probe can then be calibrated or adjusted by comparison with the reference value of the humidity standard.
Applications
Calibration and adjustment of Rotronic probes (third-party probes also possible).
Order Code Salt Solution | Nominal value [%rH] | Measurement uncertainty at 23°C |
---|---|---|
EA00-SCS | 0.5 | ±0.3 %rH |
EA10-SCS | 10 | ±0.3 %rH |
EA35-SCS | 35 | ±0.4 %rH | EA80-SCS | 80 | ±0.7 %rH |