1. Product introduction:
7-inch color high-resolution (1024*600) capacitive touch screen, high sensitivity, complete parameter display, intuitive, convenient and simple to use. It meets the GLP requirements of international standards, and has intelligent functions such as automatic calibration, automatic temperature compensation, data storage, USB output, clock display, wireless printing, function settings and self-diagnostic information. The instrument has a built-in Chinese and English bilingual operating system, humanized operation design, the instrument's brightness is adjustable, and the built-in large-capacity storage can store 2000 sets of measurement data. The data can be stored and transferred through a U disk. The stored data can be opened by the general office software EXELL, which is convenient for customers to edit.
二、技術指標:
◆測量范圍:pH:(0~14. 00)pH105) ℃ mV: (0~±1999)mV T: (-5-105)℃
◆示值誤差:pH:±0. 01pH mV: ±0. 1%F. S T: ±0. 5℃
◆分辨率:pH:0. 01pH mV: 1mV T: 0. 1℃
◆輸入阻抗:≥1x10Ω
◆穩(wěn)定性:≤0. 01pH±1個字/2小時
◆溶液溫度補償范圍:0-100℃
◆外形尺寸:210x78x40mm
◆重量:0. 2kg
◆消耗功率:7m W
2. Technical indicators:
◆Measuring range: pH: (0~14. 00)pH105) ℃ mV: (0~±1999)mV T: (-5-105)℃
◆Indication error: pH: ±0. 01pH mV: ±0. 1%F. S T: ±0. 5℃
◆Resolution: pH:0. 01pH mV: 1mV T: 0. 1℃
◆Input impedance: ≥1x10Ω
◆Stability: ≤0.01pH±1 word/2 hours
◆Solution temperature compensation range: 0-100℃
◆Dimensions: 210x78x40mm
◆Weight: 0.2kg
◆Power consumption: 7m W
三、工作原理:
水溶液pH值的測量一般用玻璃電極作為指示電極,*電極作為參比電極,當溶液中氫離子濃度(嚴格說是活度)即溶液的pH值發(fā)生變化時,玻璃電極和*電極之間產(chǎn)生的電勢也隨著發(fā)生變化,而電勢變化關系符合下列公式:
ΔE=-58. 16xΔpHx (273+t℃) /293 (mV)
ΔE-表示電勢的變化,以毫伏為單位。(mV)
ΔpH-表示溶液pH值的變化。
t-表示被測溶液的溫度(℃).
常用的指示電極有玻璃電極、銻電極、氟電極,銀電極等等,其中玻璃電極
使用廣。pH玻璃電極頭部是由特殊的敏感薄膜制成,它對氫離子有敏感作用,它插入被測溶液內(nèi),其電位隨被測液中氫離子的濃度和溫度而改變。在溶液溫度為25℃時,每變化1個pH,電極電位就改變59. 16mV.這就是常說的電極的理論斜率系數(shù)。常用的參比電極為*電極,其電位不隨被測液中氫離子濃度而改變。pH值測量的實質(zhì)就是測量兩電極間的電位差。當一對電極在溶液中產(chǎn)生的電位差等于零時,被測溶液的pH值即為零電位pH值,它與玻璃電極內(nèi)溶液有關。本儀器配用的是由玻璃電極和Ag-Agcl電極組成一體的復合電極,其零點位pH值在7±0. 25pH.
3. Working principle:
The measurement of pH value of aqueous solution generally uses a glass electrode as an indicator electrode, and a calomel electrode as a reference electrode. When the hydrogen ion concentration (strictly speaking, the activity) in the solution, that is, the pH value of the solution, changes, the glass electrode and the calomel electrode The electric potential generated during the time also changes, and the electric potential change relationship conforms to the following formula:
ΔE=-58. 16xΔpHx (273+t℃) /293 (mV)
ΔE- represents the change in electric potential, in millivolts. (MV)
ΔpH- indicates the change in pH value of the solution.
t-represents the temperature of the tested solution (℃).
Commonly used indicator electrodes include glass electrodes, antimony electrodes, fluorine electrodes, silver electrodes, etc among which glass electrodes
Widely used. The pH glass electrode head is made of a special sensitive film, which is sensitive to hydrogen ions. It is inserted into the test solution, and its potential changes with the concentration and temperature of the hydrogen ions in the test solution. When the solution temperature is 25°C, the electrode potential changes 59.16mV for every change of pH. This is the theoretical slope coefficient of the electrode that is often said. The commonly used reference electrode is calomel electrode, and its potential does not change with the concentration of hydrogen ions in the test solution. The essence of pH measurement is to measure the potential difference between two electrodes. When the potential difference produced by a pair of electrodes in the solution is equal to zero, the pH value of the measured solution is the zero potential pH value, which is related to the solution in the glass electrode. This instrument is equipped with a composite electrode composed of a glass electrode and an Ag-Agcl electrode, and its zero point pH value is 7±0. 25pH.
四、測定樣品的注意事項:
◆儀器的電極插頭和插口必須保持清潔干燥,不使用時應將短路插頭或電極
插頭插上,以防止灰塵及濕氣浸入而降低儀器的輸入阻抗,影響測定準確性。
◆不同的樣品,應選擇相適應的pH電極(例如:測量強酸、強堿或者純水等)。
◆在樣品測量時,電極的引入導線須保持靜止,不要用手觸摸。否則將會引
心起測量不穩(wěn)定。
◆配制標準溶液必須使用二次蒸餾水或去離子水,其電導率應小于2μS/cm,
*好煮沸使用。
◆要保證標準緩沖液的準確可靠,堿性溶液應裝在聚乙烯瓶中密封蓋緊。標
準緩沖液應存放在冰箱(低溫5~10℃)中保存,一般可保存2-3個月。如發(fā)現(xiàn)有
渾濁、發(fā)霉、或沉淀等現(xiàn)象時,不能繼續(xù)使用。
◆勿使用超過保質(zhì)期的標準緩沖液,勿將使用過的標準緩沖液倒回標準液儲
藏瓶中。
◆標定時,盡可能用接近樣品pH 值的標準緩沖液進行標定,且樣品的溫度
盡可能與標定液的溫度一致。
◆精密測量時,建議使用ATC溫度探頭。
◆在儀器使用過程中若更換電極,重新進行標定。
Fourth, the matters needing attention in the determination of samples:
◆The electrode plug and socket of the instrument must be kept clean and dry, and the plug or electrode should be short-circuited when not in use.
Plug in the plug to prevent dust and moisture from entering and reducing the input impedance of the instrument, which affects the accuracy of the measurement.
◆Different samples should choose a suitable pH electrode (for example: measuring strong acid, strong alkali or pure water, etc.).
◆When measuring the sample, the lead-in lead of the electrode must be kept still, and do not touch it with your hands. Otherwise it will lead to
Heart measurement is unstable.
◆Double distilled water or deionized water must be used to prepare standard solution, and its conductivity should be less than 2μS/cm,
It is best to boil and use.
◆To ensure the accuracy and reliability of the standard buffer solution, the alkaline solution should be sealed in a polyethylene bottle and tightly capped. Mark
The quasi-buffer should be stored in the refrigerator (low temperature 5~10℃), generally can be stored for 2-3 months. If found
Do not continue to use it when it becomes turbid, moldy, or precipitates.
◆Do not use the standard buffer solution beyond the shelf life, and do not pour the used standard buffer solution back into the standard solution storage
Hidden in a bottle.
◆When calibrating, use standard buffer as close as possible to the pH value of the sample for calibration, and the temperature of the sample
It should be as consistent as possible with the temperature of the calibration solution.
◆For precise measurement, it is recommended to use ATC temperature probe.
◆If the electrode is replaced during the use of the instrument, re-calibrate.