代碼 | 275顯示信息 | 可能的故障 |
A1 | EEPROM Checksum - Core Processor | 流量計需要返修
|
A2 | RAM Error - Core Processor | 重新上電/流量計需要返修 |
A3 | Sensor failure | 檢查測試點 |
A4 | Temperature out of range | 檢查測試點 |
A5 | Input over range | 檢查測試點
|
A6 | Field device not characterizeded | 檢查特性化. 特別要檢查FCF和K1值 |
A7 | Real time interrupt failure | 流量計需要返修
|
A8 | Density outside limits | 檢查測試點 |
A9 | Field device warming up | 流量計預(yù)熱
|
A10 | Calibration failed | 重新上電, 然后重新標(biāo)定 |
A11 | Excess calibration correction, zero too low | 重新上電, 然后重新調(diào)零
|
A12 | Excess calibration correction, zero too high | 重新上電, 然后重新調(diào)零 |
A13 | Process too noisy to perform auto zero | 消除或降低干擾, 重新調(diào)零 |
A14 | Electronics failure | 變送器需要返修 |
A15 | Data loss possible | 重新上電,檢查組態(tài)數(shù)據(jù) |
A16 | Line RTD Overrange | 檢查測試點 |
A17 | Case RTD Overrange | 檢查測試點
|
A18 | EEPROM Checksum - 1000/2000 | 變送器需要返修 |