油浸式試驗變壓器\高壓試驗變壓器型號含義:
油浸式試驗變壓器\高壓試驗變壓器技術(shù)參數(shù):
型號規(guī)格 | 容量KVA | 低壓輸入 | 交流高壓輸出 | 直流高壓輸出 | 測量 | ||
電壓 | 電流 | 電壓 | 電流 | ||||
YDJ-1.5/50 | 1.5 | 200 | 7.5 | 50 | 30 | 70 | 500 |
YDJ-3/50 | 3 | 200 | 15 | 50 | 60 | 70 | 500 |
YDJ-5/50 | 5 | 200 | 25 | 50 | 100 | 70 | 500 |
YDJ-10/50 | 10 | 200 | 50 | 50 | 200 | 70 | 500 |
YDJ-20/50 | 20 | 400 | 28 | 50 | 400 | 70 | 500 |
YDJ-30/50 | 30 | 400 | 43 | 50 | 600 | 70 | 500 |
YDJ-50/50 | 50 | 400 | 72 | 50 | 1000 | 70 | 500 |
YDJ-10/100 | 10 | 200 | 50 | 100 | 100 | 140 | 1000 |
YDJ-20/100 | 20 | 400 | 28 | 100 | 200 | 140 | 1000 |
YDJ-30/100 | 30 | 400 | 43 | 100 | 300 | 140 | 1000 |
YDJ-50/100 | 50 | 400 | 72 | 100 | 500 | 140 | 1000 |
YDJ-100/100 | 100 | 400 | 144 | 100 | 1000 | 140 | 1000 |
YDJ-20/150 | 20 | 400 | 28 | 150 | 133 | 210 | 1000 |
YDJ-30/150 | 30 | 400 | 43 | 150 | 200 | 210 | 1000 |
YDJ-50/150 | 50 | 400 | 72 | 150 | 333 | 210 | 1000 |
YDJ-100/150 | 100 | 400 | 144 | 150 | 666 | 210 | 1000 |
YDJ-150/150 | 150 | 400 | 216 | 150 | 1000 | 210 | 1000 |
YDJ-50/200 | 50 | 400 | 72 | 200 | 250 |
| 1000 |
YDJ-100/200 | 100 | 400 | 144 | 200 | 500 |
| 1000 |
操作步驟:
1、按接線原理圖連接好引線,并將變壓器和控制箱可靠接地;
2、試驗前,檢查各部位的接線是否接觸良好,并檢查控制箱的調(diào)壓器是否調(diào)至"零"位;
3、接通電源,綠色指示燈亮,按下啟動按鈕,紅色指示燈亮,表示變壓器已通電,等待升壓;
4、順時針勻速旋轉(zhuǎn)調(diào)壓器手柄,進行升壓,并密切注意儀表指示以及試品的情況;
5、試驗完畢后,應(yīng)迅速將電壓降至零位,并按下停止按鈕,然后切斷電源,解開試驗引線。
注意事項:
1、做高壓電氣試驗時,必須由2人或2個以上人員參加,并明確做好分工,明確相互間的方法。并有專人監(jiān)護現(xiàn)場安全及觀察試品的試驗狀態(tài);
2、高壓試驗變壓器和試驗變壓器控制箱應(yīng)有可靠的接地;
3、試驗過程中,升壓速度不能太快,也決不允許突然全電壓通電或斷電;
4、在升壓或耐壓試驗過程中,如發(fā)現(xiàn)下列不正常情況時,應(yīng)立即降壓,并切斷電源,停止試驗,查明原因后再做試驗。① 電壓表指針擺動很大;② 發(fā)現(xiàn)絕緣燒焦的異味、冒煙現(xiàn)象;③ 被測試品內(nèi)有不正常的聲音。
5、試驗中,如果試品短路或故障擊穿,控制箱中的過流繼電器工作,此時,將調(diào)壓器回至零位,并切斷電源后,方可將試品取出。
6、進行電容試驗或進行直流高壓泄漏試驗時,試驗完畢后,將調(diào)壓器降至零位后,并切斷電源,然后,應(yīng)用放電棒將試品或電容器的高壓端對地進行放電,以免存留在電容中的電勢發(fā)生觸電危險。
誠摯邀請各地有信譽的經(jīng)銷商代理我們的產(chǎn)品,歡迎各用戶單位來我們公司指導(dǎo)工作,我們愿意共同努力,為實現(xiàn)供需雙贏盡自己的一份力量!訂購:,服務(wù)永遠(yuǎn)伴隨您。