5EH-505S RF admittance meter meter RF admittance measurement principle of levelmeasurement, shielding its unique technology, a fully solve the materials andconditions linked to the probe temperature and pressure changes caused by drift andmalfunction, instrument Work is extremely stable and reliable.
RF admittance meter from the probe, electronic unit, protective shell composition, can be used whole or split installation. High sensitivity, good stability, is not linked to the significant characteristics of materials affect can be used for a variety of liquids,pastes, solid particles or powder of the limit control, the output for the DPDT relay contact signal.
Features:
Maintenance-free: no mechanical wear parts, no regular maintenance or replacement
Anti-linked material: the design of the circuit shielding technology, so that instrumentslinked to ignore the material of the probe, without regular cleaning or repeatedcalibration
Wide range of applications: to measure liquid, paste, solid particles and powder
Easy installation: can be threaded through the mouth or on the sink flange for installation, you can choose whole or split installation, commissioning, simple and convenient.
Technical parameters
Power supply: AC 185V-256V 2W or DC 15V-30V 2W
Output: DPDT relay switch signal (which can be set high action or low-site action)
Contact capacity: AC 220V 5A non-inductive, 3A felt
Sensitivity: 0.3PF Accuracy: ± 0.5%
Repeatability: conductive materials, 1.6mm, 50mm insulation materials
Load Resistance: Center-ended to 150Ω, center-ended shielded 500Ω, 150Ωshielded end to ground
Ambient temperature: -40 ℃ -80 ℃ Probe material: 1Cr18Ni9Ti
Probe Length: 500mm (standard)> 50mm 1000mm
Split cable: single double shielded cable, length 5 m (standard), up to 50 meters
Process connection: A3/4NPT thread installation (standard) or optional flange H
Housing protection: IP65 protection standards consistent with
Explosion-proof grade: Exd Ⅱ BT4
Working temperature and working pressure: (live: special type, can be special)