T2 Series T2RC/T2C |
The Tenney T2 temperature and humidity cycling chamber simulates a wide range of temperature and humidity conditions while utilizing minimal floor space. We specifically designed these environmental chambers to have a compact exterior and an ample interior workspace.
The T2 Temperature and Humidity Cycling Chamber design allocates space within the interior of the cabinet for all standard options. This greatly improves the mobility of the unit as well as the overall appearance.
T2 Series T2RC/T2C Features: |
Vapor-tight, continuously welded stainless steel interiors |
Minimum footprint |
Units are |
Touch screen controller provides precise control over chamber operations and monitoring, and is mounted at eye level (VT III and Watlow F4 are not touch screen) |
Low workspace height to ease the loading and unloading of product |
Accurate control and display with minimum maintenance |
T2 Series T2RC/T2C Specifications: |
Model | T2RC | T2C | |
Workspace inches-(mm) | W | 16 (406) | 16 (406) |
D | 13 (330) | 13 (330) | |
H | 15 (381) | 15 (381) | |
Exterior inches-(mm) | W | 26 (660) | 27 (685) |
D | 26 (660) | 27 (685) | |
H | 67 (1702) | 65 (1651) | |
Temperature Range & Humidity | |||
Low | °C | -75 | -75 |
High | °C | +200 | +200 |
Change Rates in Minutes | |||
Ambient to °C Chamber Empty | 200°C | 90 Minutes | 90 |
150°C | 40 Minutes | 40 | |
100°C | 25 Minutes | 25 | |
93°C | 17 Minutes | 17 | |
65°C | 9 Minutes | 9 | |
2°C | 5 Minutes | 5 | |
-7°C | 7 Minutes | 7 | |
-12°C | 10 Minutes | 10 | |
-40°C | 30 Minutes | 30 | |
-54°C | 45 Minutes | 45 | |
-65°C | 70 Minutes | 70 | |
-73°C | 100 Minutes | 100 | |
-75°C | Ultimate | Ultimate | |
Utilities, etc . | |||
Refrigeration | (2)1/2 HP + 1/2 HP | (2)1/2 HP + 1/2 HP | |
Heater Capacity | 500 W | 500 W | |
Humidifier | W | 500 | - |
GPH | 0 .15 | - | |
AMPS @ 115V, 1Ph | 18 | 18 | |
AMPS Fuse | 20 | 20 | |
Unit Weight | Lbs/Kg | 350/158 | 350/158 |