Load section ELA-305ELB-305ELC-305CurrentVoltageInternal min resistance(*9)Internal inductance(Typ)(*9)Min operating voltage(*8, 9)Peak power(*7)20us or less2s or less20s or lessRated powerCurrent range
(Setting range)HMLNominal resolution(*1)HMLAccuracy(*4)HMLVoltage range20V50V85VCurrent range
(Setting range)Current range:HCurrent range:MNominal resolution(*1)Current range:HCurrent range:M Voltage rangeCurrent range
(Setting range)Current range:HCurrent range:MNominal resolution(*1)Current range:HCurrent range:MAccuracy(*2, 4, 6)Voltage range
(Setting range)HLNominal resolution(*1)HLAccuracyResponse timePower rangeCurrent range:HCurrent range:MNominal resolution(*1)Current range:HCurrent range:MAccuracy(*4)Current range
(Setting range)HMNominal resolution(*1)HMAccuracy(*4, 5)Control voltageControl methodLoading modeSetting periodPeriod resolutionOperating modeSlew rate
(Current range)HMLMin load response time(*3)Loading modeMax stepsStep durationDuration resolutionShort current (MAX)Current range
(Setting range)HMNominal resolution(*1)HM
Max rating | ||||
60A | 30A | 12A | ||
120V | 300V | 500V | ||
17m ohm | 67m ohm | 250m ohm | ||
250nH | 400nH | 2uH | ||
1V(60A) / 0.5V(30A) / 0.2V(12A) | 2V(30A) / 1V(15A) / 0.5V(7.5A) | 3V(12A) / 1.5V(6A) / 0.7V(2.4A) | ||
7200W | 9000W | -- | ||
800W | 800W | -- | ||
500W | 500W | -- | ||
300W | ||||
CC mode | ||||
0 - 60A | 0 - 30A | 0 - 12A | ||
0 - 6A | 0 - 3A | 0 - 1.2A | ||
0 - 0.6A | 0 - 0.3A | 0 - 0.12A | ||
5mA | 2.5mA | 1mA | ||
0.5mA | 0.25mA | 0.1mA | ||
0.1mA | 0.05mA | 0.02mA | ||
+/- 0.2% of stg. +/- 25mA+Vin/Rin | +/- 0.2% of stg. +/- 15mA+Vin/Rin | +/- 0.2% of stg. +/- 10mA+Vin/Rin | ||
+/- 0.2% of stg. +/- 12mA+Vin/Rin | +/- 0.2% of stg. +/- 6mA+Vin/Rin | +/- 0.2% of stg. +/- 3mA+Vin/Rin | ||
+/- 0.2% of stg. +/- 6mA+Vin/Rin | +/- 0.2% of stg. +/- 3mA+Vin/Rin | +/- 0.2% of stg. +/- 2mA+Vin/Rin | ||
CR mode | ||||
0.025 - 200 ohm | 0.1 - 800 ohm | 0.3 - 2.5k ohm | ||
0.25 - 2k ohm | 1 - 8k ohm | 3 - 25k ohm | ||
4ms | 1ms | 333us | ||
400us | 100us | 33us | ||
120V | 300V | 500V | ||
0.075 - 600 ohm | 0.3 - 2.5k ohm | 0.9 - 7k ohm | ||
0.75 - 6k ohm | 3 - 25k ohm | 9 - 70k ohm | ||
1.33ms | 333us | 111us | ||
133us | 33us | 11us | ||
+/- 0.5% of Conv. Curr. +/- 0.2% of f.s. +Vin/Rin | ||||
CV mode | ||||
0 - 120V | 0 - 300V | 0 - 500V | ||
0 - 20V | 0 - 50V | 0 - 85V | ||
10mV | 30mV | 50mV | ||
2mV | 5mV | 10mV | ||
+/- 0.2% of stg. +/- 0.2% of f.s. | ||||
Fast/Slow (300us/10ms Typ) | ||||
CP mode | ||||
0 - 300W | ||||
0 - 40W | ||||
50mW | ||||
5mW | ||||
+/- 2.5% of stg. +/- 1% of f.s. +/- (VinXVin)/Rin | ||||
External control mode | ||||
0 - 60A | 0 - 30A | 0 - 12A | ||
0 - 6A | 0 - 3A | 0 - 1.2A | ||
10mA | 5mA | 2mA | ||
1mA | 0.5mA | 0.2mA | ||
+/- 0.2% of stg. +/- 0.5% of f.s. +Vin/Rin | ||||
0 - 10V | ||||
Dynamic load mode | ||||
Dynamic mode | ||||
Switching | ||||
CC mode | ||||
- 20ms / - 200ms / - 2s / - 20s / - 60s | ||||
1us / 10us / 100us / 1ms / 10ms | ||||
Continuous, Single | ||||
0.2 - 20A/us | 0.1 - 10A/us | 0.01 - 1A/us | ||
0.02 - 2A/us | 0.01 - 1A/us | 0.001 - 0.1A/us | ||
0.002 - 0.2A/us | 0.001 - 0.1A/us | 0.0001 - 0.01A/us | ||
500ns | ||||
Sequence mode | ||||
CC / CR / CV / CP mode | ||||
1024 | ||||
1ms - 10min | ||||
1ms(1ms - 100ms) / 100ms(100ms - 10min) | ||||
Short mode | ||||
60A | 30A | 12A | ||
Current limit setting | ||||
0.3 - 60A | 0.15 - 30A | 0.06 - 12A | ||
0.03 - 6A | 0.015m - 3A | 0.006m - 1.2A | ||
0.3A | 0.15A | 0.06A | ||
30mA | 15mA | 6mA | ||
Parallel operation | ||||
Master-Slave operation. One master unit can control up to 4 units of slaves of same voltage rating. For example: If the master unit is ELA-305 then ELA-155, ELA-305 or ELA-1005 may be connected as slave unit. |
*1 : Nominal resolution: Approximated resolution of the range. / *2 : Conv.Curr: Ideal current value being calculated by [input / set R] / *3 : Min load response time: Minimum time required by the load when setting slew rate. / *4 : Rin: ELA-155 : 100kohm(typ) ELB-155 : 200kohm(typ) ELC-155 : 1500kohm(typ) / *5 : External controlsetting accuracy: Setting accuracy is defined when external control voltage of 10V is fed. / *6 Setting accuracy in CR mode is valid when applied 10% or more voltage of the voltage range. / *7 : It may vary in accordance to the internal temperature. / *8 Minimum operating voltage vary according to the current value. / *9 At rear loading terminals.
Measuring section ELA-305ELB-305ELC-305RangeHLResolutionHLAccuracy(*1)Measuring time(*2)Range(*3)HMLResolutionHMLAccuracy(*1)H, MLMeasuring time(*2)Measuring method(*4)Measuring time(*2)DC Voltage | ||||
120.00V | 300.00V | 500.00V | ||
20.000V | 50.000V | 85.000V | ||
10mV | 10mV | 10mV | ||
1mV | 1mV | 1mV | ||
+/- 0.05% of rdg. +/- 0.05% of f.s. | ||||
Approx. 100ms | ||||
DC Current | ||||
60A | 30A | 12A | ||
6A | 3A | 1.2A | ||
0.6A | 0.3A | 0.12A | ||
0.5mA | 0.5mA | 0.5mA | ||
0.1mA | 0.1mA | 0.1mA | ||
0.1mA | 0.1mA | 0.1mA | ||
+/- 0.2% of rdg. +/- 0.2% of f.s. | ||||
+/- 0.2% of rdg. +/- 0.5% of f.s. | ||||
Approx. 100ms | ||||
Power | ||||
Calculation by [Voltage X Current] | ||||
Approx. 200ms |
*1 Guaranteed for 6 months at 23 +/- 5C, <70% RH / *2 Measurement time when set at same measuring mode and range. / *3 Current measurement range depends on load setting range. / *※4 Absolute value of the measurement.
General ELA-305ELB-305ELC-305Loading terminalsRated AC inputPower consumptionOperating temperatureCooling systemDimensionsWeightGP-IBUSBExternal control voltageOver current protectionOver power protectionOver voltage alarmTemperature protectorReverse connection alarmGeneral | |||
on front & rear panel | |||
AC100 - 240V +/- 10% 50/60Hz | |||
< 45VA (When in AC100V) | |||
5 - 40C | |||
Forced cooling fan | |||
215(W) X 128.6(H) X 420(D) excluding projected part | |||
Approx. 10Kg | |||
Interface | |||
Conforming to IEEE488.1 | |||
Conforming to USB1.1 | |||
0 - 10V | |||
Protection, Alarm | |||
Protect load section by limitting the current. | |||
Protect load section by limitting the power. | |||
Generate alarm when over voltage is applied. | |||
Generate alarm and cut off the output when overheaded. | |||
Generate alarm when reverse voltage is applied. |
* Noted that the alarm function does not protect the instrument. Remove the cause of alarm to protect the instrument.
ELA-305ELB-305ELC-305Output(*1)Output voltageCC1CC2 - CC16Current range(*2, 3)H, MLOutput resistanceAccuracy(*4)H, MLTRIG OUT | ||||
Photo coupler output | ||||
+5V (Typ) | ||||
0V (Typ) | ||||
CURRENT MONITOR | ||||
1V / 60A f.s. | 1V / 30A f.s. | 1V / 12A f.s. | ||
0.04V / 0.6A f.s. | 0.04V / 0.3A f.s. | 0.04V / 0.12A f.s. | ||
50 ohm | ||||
+/- 1% of Conv.Volt. +/-1 % of f.s. | ||||
+/- 5% of Conv.Volt. +/- 3% of f.s. |