產(chǎn)品規(guī)格 |
Maximum Range | Voltage | ±210V |
Current | ±1.05A |
Power | 22W |
Voltage Resolution | 1µV |
Current Resolution | 10pA |
Source | DC Voltage | Output Voltage | ±21V/ ±1.05A, ±210V/±105 mA |
Current Limit | Min. 0.1% of range |
Programming Resolution | 1µV, ±200.000mV range |
10µV, ±2.00000V range |
100µV, ±20.0000V range |
1mV, ±200.000V range |
Programming Accuracy 1 | ±(0.02%+600µV), ±200.000mV range |
±(0.02%+600µV), ±2.00000V range |
±(0.02%+2.4mV), ±20.0000V range |
±(0.02%+24mV), ±200.000V range |
Load Regulation | 0.01% of range + 100µV |
Line Regulation | 0.01% of range |
Overshoot | <0.1% typical (full scale step, resistive load,10mA range) |
Recovery Time (1000%Load Change) | <250µs (within 0.1% plus load regulation errors, 1A and 100mA compliance.) |
Ripple and Noise | 4mV rms(20Hz~ 1MHz) 10mVpp(20Hz~ 1MHz) |
Temperature Coefficient | ±(0.15 × accuracy specification)/°C |
(0°–18°C & 28°–50°C) |
DC Current | Output Current | ±1.05A /±21V, ±105 mA /±210V |
Voltage Limit | Min. 0.1% of range |
Programmed Source Resolution | 10pA, ±1.00000µA range |
100pA, ±10.0000µA range |
1nA, ±100.000µA range |
10nA, ±1.00000mA range |
100nA, ±10.00000mA range |
1µA, ±100.000mA range |
10µA, ±1.00000A range |
Programmed Source Accuracy 1 | ±(0.035%+600pA), ±1.00000µA range |
±(0.033%+2nA), ±10.0000µA range |
±(0.031%+20nA), ±100.000µA range |
±(0.034%+200nA), ±1.00000mA range |
±(0.045%+2µA), ±10.00000mA range |
±(0.066%+20µA), ±100.000mA range |
±(0.27%+900µA), ±1.00000A range |
Load Regulation | 0 0.01% of range + 100pA |
Line Regulation | 0.01% of range |
Overshoot | <0.1% typical (1mA step, RL = 10kΩ, 20V range). |
Temperature Coefficient | ±(0.15 × accuracy specification)/°C |
(0°–18°C & 28°–50°C) |
General | Output Settling Time2 | 100µs typical Time |
Output Rise Time (±30%) | 300µs, 200V range, 100mA compliance. |
150V/µs, 20V range, 100mA compliance. |
DC Floating Voltage | Output can be floated up to ±250VDC |
Remote Sense | Up to 1V drop per load lead. |
Compliance Accuracy | Add 0.3% of range and ±0.02% of reading to base specification. |
Range Change Overshoot 3 | Adjacent range changes between 200mV, 2V and 20V ranges, 100mV typical. |
Minimum Compliance Value | 0.1% of range |
Command Processing Time 4 | Autorange On:10ms.Autorange Off:7ms. |