MULTIPLE RANGE CAPABILITY:8 operations ranges - 10, 30, 100, 300, 1000, 3000, 10,000, 30,000 ppm carbon
RESOLUTION: 0.01 ppm Carbon (lowest range)
REPEATABILITY: Better than � 0.5% of full scale
LINEARITY: Better than 1% of full scale
O2 EFFECT: Less than 2% of full scale
CH4 EFFECT: Less than 1.3 times propane
RESPONSE TIME: 90% of full scale in 1.5 seconds
SAMPLE FLOW REQUIREMENT: 4.0 liter/min. �1.5 liter/min.
NOISE: Less than � 0.5% of full scale
ZERO & SPAN DRIFT: Less than 1% of full scale per 24 hours
FLOW CONTROL: Electronic Proportional Pressure Controller
FUEL REQUIREMENTS: 40% H2/60% He (100cc/min.) or 99% H2 (30cc/min.) (specify at time of order) Fuel inlet pressure 25 psig
READOUT: As ppm CH4 or C3H8 (specify at time of order)
AIR REQUIREMENTS: Less than 1 ppm Carbon Purified or synthetic air (200cc/min.) Air Inlet Pressure 25psig
DISPLAY: 31/2 Digit Panel Meter
DIAGNOSTICS: 3-1/2 Digital Meter with 7 Position Switch, Collector Voltage 15 VDC, Fuel Pressure, Burner Temperature, Air Pressure, Catalyst Temperature, Sample Pressure
ANALOG OUTPUT: 0-10 VDC & 4-20 mA DC/0-20 mA DC
FUEL/AIR CONTROL: Forward Pressure Regulator
IGNITION: Momentary push-button with Flame-On Indicator (manual or remote control)
FLAME OUT INDICATOR: With automatic fuel shut-off and dry contacts at rear panel
AMBIENT TEMPERATURE: 5-45�C
SAMPLE TEMPERATURE RANGE: 0-50�C
WARM-UP TIME: 1 hour
FITTINGS: 1/4� tube
POWER REQUIREMENTS: 115/230 (�10%) VAC @ 50/60 Hz, 40 watts
DIMENSIONS: 5-1/4�H x 19� W x 23� D
(133mm x 483mm x 508mm)
RELATIVE HUMIDITY: Less than 90% RH non-condensing
WEIGHT: 28 lbs. (12.3 kg)
Specifications subject to change without notice.
sample/fuel/air pressures (no gauges)
The Methane Only Option provides the capability for measuring METHANE ONLY or TOTAL HYDROCARBONS. The choice is switch selectable from the front panel or by remote contact closure. Additional hardware includes a heated catalyst assembly and a three-way valve. Conversion efficiency is greater than 98% for propane and above. (This option requires the use of 40/60% H2/He fuel and a readout as ppm CH4).