參數(shù)SPECIFICATIONS
DETECTOR: Flame Ionization Detector (FID)
CH4/THC USER DEFINABLE RANGES:0-3 ppm to 3% C
RESOLUTION: 10 ppb Carbon (lowest range)
REPEATABILITY: Better than ± 0.5% of full scale
LINEARITY: Better than ± 0.5% of full scale
O2 EFFECT: Less than 2% of full scale H2/He Fuel
CH4 EFFECT: Less than 1.15 times Propane
RESPONSE TIME: T90 < 1.0 Secs. to 60 Secs.
SAMPLE FLOW RATE:1.5 to 3.0 LPM
SAMPLE TEMPERATURE: Up to 191°C, Non-Condensing (HFID), 85°C Non-Condensing (FID)
OVEN TEMPERATURE: 200°C HFID (85°C FID)
INTERNAL SAMPLE FILTER: 0.1 micron replaceable filter provided
NOISE: Less than ± 1% of full scale
ZERO & SPAN DRIFT: Less than 1% of full scale per 24 hours
FLOW CONTROL: Electronic Proportional Pressure Controller
FUEL REQUIREMENTS: 40% H2/60% He (100cc/min.) or 99% H2 (30cc/min.) (specify at time of order) Fuel inlet pressure 25 psig
READOUT: As ppm CH4 or C3H8 (specify at time of order)
AIR REQUIREMENTS: Less than 1 ppm Carbon
Purified or synthetic air (200cc/min.) Air Inlet Pressure 25psig
DISPLAY: 3" x 5" Back Lit LCD
DIAGNOSTICS: Oven Temperature, Burner Temperature, Cutter Temperature, Sample/Fuel/Air Pressure, Flow Rates and EPC Control Voltages
KEYPAD DISPLAYS: Factory Settings, TCP/IP address, Passwords(4), Scalable Analog Output Voltages, Full Scale Range Select, Auto Cal Times
SPECIAL FEATURES: Calculated NMHC, Auto Ranging, Auto Calibration (Adjustable through internal clock)
ANALOG OUTPUT: 0-10 VDC & 4-20 mA DC/0-20 mA DC
FUEL/AIR CONTROL: Electronic Proportional Pressure Controller
IGNITION: local, remote or automatic
AMBIENT TEMPERATURE: 5-45°C
WARM-UP TIME: 1 hour
FITTINGS: 1/4â tube
POWER REQUIREMENTS: 115/230 (±10%)VAC@50/60Hz,750watt
DIMENSIONS: 5-1/4âH x 19âW x 23âD
(133mm x 483mm x 508mm)
WEIGHT: 50 lbs.
RELATIVE HUMIDITY: Less than 90%
特點(diǎn)FEATURES
· CE Approved
· Made in USA
· Sensitivity 0.01 ppm
· Automatic ignite capability - local or remote
· Automatic fuel/air shut off
· Automation Ranging/ Calibration
· Temperature Stabilized Detector
· Internal heated filter with external access
· Calculates, Displays, & Outputs CH4/NMHC/THC
· Electronic proportional pressure sample controller
· Multiple User Definable Ranges 0-3 to 3%
· Optional continuous methane only
· Digital diagnostics
· 19" rack mount (only 5-1/4" high)
應(yīng)用APPLICATIONS
· EPA Method 25A compliance monitoring
· Stack gases (CEM / MACT / Process)
· Vehicle emissions - Diesel
· VOC Abatement/ Scrubber Efficiency
· Carbon bed breakthrough detection
· Fermentation Monitoring
· Personnel Safety
· Fenceline (perimeter) monitoring
· Non-methane hydrocarbon measurements
其他:
DESCRIPTION
The CAI Model 600 FID/HFID Analyzers are designed to continuously measure the total concentration of hydrocarbons within a gaseous sample. The analyzers exhibit superior sensitivity and response time. The gaseous sample can be exhaust gases from an internal combustion engine, a combustion process or VOC abatement systems. The instrument can measure THC in the ppm or percent levels. All internal components are maintained at the oven set temperature preventing condensation. Options for Internal Heated Sample Pump and Methane Only measurements are available.
METHOD OF OPERATION
The Model 600 FID/HFID utilizes the principle of Flame Ionization Detection (FID) to determine the total hydrocarbon concentration within a gaseous sample. The HFID analyzer has a heated oven (191°C)which contains a burner and an optional heated pump. The small flame of the burner is elevated and sustained by regulated flows of air and either pure hydrogen or a 40/60 mixture of hydrogen and (helium or nitrogen). The split ring detector contains 2 electrodes. One electrode is negatively polarized using a precision power supply and the other electrode, known as the "collector" is connected to a high impedance, low noise electronic amplifier. The two electrodes establish an electrostatic field. When gaseous sample is introduced to the burner, it is ionized in the flame and the electrostatic field causes a small current bewteen the electrodes. This current is measured by the precision electrometer amplifier and is directly proportional to the hydrocarbon concentration of the sample.
.
METHANE ONLY OPTION
The Methane Only Option provides the capability for measuring METHANE ONLY or TOTAL HYDROCARBONS and calculates, displays and outputs the sample's NMHC content. Conversion efficiency is greater than 98% for Ethane and heavier hydrocarbons. ( This option requires the use of 40/60% H2/He fuel and a readout as ppm CH4 ).