5LC 二氧化硫檢測(cè)管 Sulphur dioxide
Sulphur dioxide No.5LC
SO2
Performance |
Measuring Range | 0.1 to 0.25 ppm | (0.25) to 10 ppm | 10 to 25 ppm |
Number of Pump Strokes | 4 | 2 | 1 |
Correction Factor | 0.4 | 1 | 2.5 |
Sampling Time | 2 minutes per pumpstroke | ||
Detecting Limit | 0.02 ppm (n=4) | ||
Color Change | Bluish purple →White | ||
Reaction Principle | SO2 +I2 +2H2O 仺 H2SO4 + 2HCl | ||
Coefficient ofVariation | 10% (for0.25 to 2 ppm), 5% (for 2 to 10 ppm) | ||
ShelfLife | 3Years | ||
Correctionsfor temperature & humidity | Temperature correction is necessary | ||
Store the tubes at cool and darkplace. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Nitrogendioxide | - | Minus error | Produces brown stainof 1 mm |
Hydrogensulfide | 1/10 | Plus error | Produce uncleardemarcation |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | Permeation tubemethod |
TLV-TWA | TLV-STEL | Explosive range |
2ppm | 5ppm | - |