5H二氧化硫檢測(cè)管 Sulphur dioxide
Sulphur dioxide No.5H
SO2
Performance |
MeasuringRange | 0.05 to 0.5% | 0.5 to 4.0% | 4.0 to 8.0% |
Number of PumpStrokes | 10 to 2 | 1 | 1/2 |
CorrectionFactor | 1/2 to 1/10 | 1 | 2 |
SamplingTime | 1 minute per pumpstroke | ||
DetectingLimit | 0.01% (n=10) | ||
ColorChange | Orange → Green | ||
ReactionPrinciple | SO2 + Cr6+→ Cr3+ | ||
Coefficient ofVariation | 10% (for 0.5 to 1 %),5% (for 1 to 4 %) | ||
ShelfLife | 3Years | ||
Correctionsfor temperature & humidity | Unnecessary | ||
Store the tubes at cool and darkplace. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Carbonmonoxide | - | No effect | Nodiscoloration |
Carbondioxide | - | No effect | Nodiscoloration |
Nitrogendioxide | - | No effect | Nodiscoloration |
Hydrogensulfide | 500ppm | Plus error | Nodiscoloration |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | High pressure gascylinder method |
TLV-TWA | TLV-STEL | Explosive range |
2ppm | 5ppm | - |