pl2210系列二極管泵浦皮秒Nd:YAG激光器
產品簡介:
短脈寬、高能量,高達2 kHz的重復率和優(yōu)良的脈沖穩(wěn)定性進行pl2210系列二極管泵浦皮秒激光器的許多應用的,如材料加工、激光光譜學、光學參量發(fā)生器泵和其他任務。
產品特點:
- 高脈沖能量
- 二極管泵浦固體設計
- 空氣冷卻,不需要外部供水
- 交鑰匙工程
- 維護成本低
- 可選條紋相機觸發(fā)脈沖與< 10的均方根抖動
- 遙控板
- 通過USB接口與PC機提供的控制
- LabVIEW? drivers
- 可選的溫度穩(wěn)定性,第二,第三,第四諧波發(fā)生器
產品應用:
- 時間分辨熒光,泵浦-探測光譜
- OPG/OPA/OPO泵浦
- 其他光譜和非線性光學中的應用
- 激光遙感
Specifications 1)
Model | PL2210 | PL2210A | PL2210B | PL2210B-TR | PL2210C | |
Output energy | ||||||
at 1064 nm | 0.4 mJ | 0.9 mJ | 2.5 mJ | 2.5 mJ at 1 kHz | 2.5 mJ / 5 mJ | |
at 532 nm2) | 0.2 mJ | 0.45 mJ | 1.3 mJ | - | 1.3 mJ / 2.5 mJ | |
at 355 nm3) | 0.11 mJ | 0.3 mJ | 0.8 mJ | - | 0.8 mJ / 1.3 mJ | |
at 266 nm4) | 0.05 mJ | 0.15 mJ | 0.5 mJ | - | 0.5 mJ / 0.8 mJ | |
Pulse energy stability (Std. Dev)5) | ||||||
at 1064 nm | 0.5 % | |||||
at 532 nm | 0.8 % | #colspan# | ||||
at 355 nm | 1 % | |||||
at 266 nm | 2.5 % | |||||
Pulse duration (FWHM) 6) | 25±2 ps | 25±2 ps | 80±8 ps | 100±10 ps | 25±2 ps / 80±8 ps | |
Pulse duration stability 7) | ±1 ps | ±1 ps | ±3 ps | ±3 ps | ±1 ps / ±3 ps | |
Pulse repetition rate 8) | 1 or 2 kHz | 1 kHz | 1 kHz / 1 MHz | 1 kHz | ||
Triggering mode | internal/external | |||||
Typical SYNC OUT pulse delay9) 10) | -500...50 ns | |||||
SYNC OUT pulse jitter (Std. Dev) | <0.1 ns rms | |||||
Spatial mode11) | TEM00 | |||||
Beam divergence12) | <1.6 mrad | |||||
Beam diameter 13) | ~3 mm | |||||
Beam pointing stability 14) | <30 μrad | |||||
Pre-pulse contrast | >200:1 | |||||
Polarization | linear, >100:1 | |||||
Physical characteristics | ||||||
Laser head size (W x L x H), mm 15) | 455 x 1035 x 242 | 455x1235x242 mm | ||||
Power supply size (W x L x H), mm | 365 x392 x 290 | 475 x 460 x 290 | 365x285x360 mm | |||
Operating requirements | ||||||
Water service | not required, air cooled | |||||
Relative humidity | 10-80 % (non condensing) | |||||
Ambient temperature | 22 ± 2 °C | |||||
power requirements | 100-240 VAC, single phase, 50/60 Hz | |||||
Power consumption 16) | <0.5 kW | <1 kW | <1 kW | <2 kW | <1.5 kW | |
1) Due to continuous improvement, all specifications are subject to change without notice. Parameters marked typical are not specifications. They are indications of typical performance and will vary with each unit we manufacture. Unless stated otherwise, all specifications are measured at 1064 nm.
2) For PL2210x-SH and PL22210x-SH/FH options. Outputs are not simultaneous. Please inquire for pulse energies for other harmonic generator options.
3) For PL2210x-TH option. Outputs are not simultaneous. Please inquire for pulse energies for other harmonic generator options.
4) For PL2210x-SH/FH option. Outputs are not simultaneous. Please inquire for pulse energies for other harmonic generator options.
5) Averaged from 300 pulses at 1 kHz pulse repetition rate.
6) Optional 80±8 ps duration.
7) Measured over 1 hour period when ambient temperature variation is less than ±2 °C.
8) Should be specified when ordering. Inquire for custom pulse repetition rates.
9) In respect to optical pulse. <10 ps rms jitter is provided with PRETRIG option.
10) SYNC OUT lead or delay can be adjusted with 0.25 ns steps in specified range.
11) Gaussian fit >90%.
12) Full angle measured at the 1/e2 point at 1064 nm .
13) Beam diameter is measured at 1064 nm at the 1/e2 point.
14) Rms value measured from 300 shots.
15) 455×1235×250 mm (W×L×H) laser head size might be required for some optional configurations.
16) At 1 kHz pulse repetition rate.
Options
- Option PRETRIG provides low jitter pulse for streak camera triggering with lead/delay in -150...150 μs range and <10 ps rms jitter.
- Option P80 provides 80±8 ps output pulse duration. Main specifications:
Model | PL2210 | PL2210A |
Pulse energy 1) | 0.7 mJ | 1.2 mJ |
Pulse duration (FWHM) | <80 ps |
*At 1064 nm
- Option PC allows reduction of the pulse repetition rate of the PL2210 series laser by integer numbers. Single shot mode is also possible. In addition, the –PC option reduces the low-intensity quasi-CW background that is present at laser output at 1064 nm wavelength. Please note that the output of fundamental wavelength and harmonics will be reduced by approx. 20% with installation of the –PC option.