Body voltage measurement test set-up. The software "Digilloscope" can mark the five highest peaks automatically and calculates the average value.
1. WT5000 - Electrometer for body voltage measurement with a handheld electrode. It has an adjustable audible alarm signal and the following parameters: It has an adjustable audible alarm signal and the following parameters:
- Test range: ± 5000 volts / ± 500 volts
- Accuracy: 5%
- Input impedance: > 10^14 Ohm
- Input capacity: < 5 pF
- Response time: <= 10 ms
- Display: Logarithmic LED-display ± 500 V, changeable to ± 5000 V with Peak Hold function
- Output: Analog ±5V DC for each range
- Power supply: Rechargeable batteries
- Complies with IEC 61340-4-5
2. A/D-Converter:
- Input impedance: 10 MOhm
- Interface: USB 2.0 / USB 1.1
- Input voltage: ± 5V
- Resolution 12 Bit
- Max. sampling rate 10 kHz
- Software Digilloscope for WINDOWS®
- Power supply via USB port
- WT5000 Electrometer
- A/D-converter with software Digilloscope
- rechargeable batteries
- power supply
- handheld electrode
- connecting cables
- conductive carrying case
- user's manual
- calibration certificate